Features
Detect document; record and compare leakage current readings over time as a means of preventing unplanned downtime; and identifying intermittent GFCI and RCD trips; all without taking equipment off line.
- True-rms measurements for accuracy when measuring complex; non-sinusoidal waveforms
- 61 mm jaw opening
- Highest resolution of 1 µA; measure up to 60 A
- Selectable filter function removes unwanted noise
- Max/Min/Average readings and hold function
- Forward-facing LED worklight for use in dark wiring cabinets
- Backlit display; auto backlight off and auto power off for extended battery life
- CAT III 600V safety rating
What's Included?
- Leakage Clamp Meter
- Soft case
- Safety manual
- Batteries: 2 AA
Specifications
-
Electrical Specifications |
Measurement function |
Alternating current |
AC ampere range |
3 mA, 30 mA, 300 mA, 3 A, 30 A, 60 A |
Range selection |
mA / A: manual selection |
3 mA / 30 mA / 300 mA: automatic selection |
3 A / 30 A / 60 A: automatic selection |
Range / Resolution |
3 mA⊃1; / 0.001 mA |
30 mA / 0.01 mA |
300 mA / 0.1 mA |
3 A / 0.001 A |
30 A / 0.01 A |
60 A / 0.1 A |
Accuracy – filter activated (40 to 70 Hz), filter off (40 to 1 kHz)⊃2; |
3 mA – 30 A |
1% + 5 digits |
60 A |
2% + 5 digits |
Frequency |
40 Hz to 1 kHz |
Crest Factor |
3 |
After calibration accuracy is valid for one year. Accuracy is expressed as ± (% of reading + digits). Reference conditions 23 ±5 °C and a maximum relative humidity of 80%. |
⊃1;The minimum value is 10 μA rms. |
⊃2;Outside of the TC / °C 18 °C to 28 °C, 0.02 + 1 |
Physical |
Display (LCD) |
Digital readout: 3300 count |
Display refresh rate |
4 times / sec |
Maximum conductor diameter |
61 mm |
Dimensions |
257 x 116 x 46 mm |
Weight |
600 g |
Battery |
2 AA, IEC LR6, NEDA 15A, alkaline |
Battery life |
Expected battery life, without using backlight and spotlight, is more than 150 hours |
Automatic shut-down |
Meter automatically shuts down after 15 minutes of inactivity |
Safety and Environmental Specifications |
General safety |
IEC 61010-1: Pollution Degree 2 |
Measurement safety |
IEC 61010-2-032: CAT III 600 V / CAT IV 300 V |
Operating temperature |
-10 °C to +50 °C |
Storage temperature |
-40 °C to +60 °C |
Operating humidity |
Non-condensing (<10> |
90% relative humidity (10 °C to 30 °C) |
75% relative humidity (30 °C to 40 °C) |
45% relative humidity (40 °C to 50 °C) |
Protection class |
IEC 60529: IP30 (jaw closed) |
Operating altitude |
2000 m |
Storage altitude |
12000 m |
Current sensor action category |
IEC 61557-13: Class 1, ≤100 A / m |
Electromagnetic compatibility (EMC) |
International |
IEC 61326-1 |
Industrial Electromagnetic Environment |
CISPR 11 |
Group 1 Class B |
Group 1 |
Generated inside the equipment and / or use radio frequency energy associated with conducting the energy for the device’s own internal functions is essential. |
Class B |
Equipment for home appliances and residential buildings directly connected to a low voltage power supply network equipment. When this device is connected to a test object, it may exceed the emission levels produced by CISPR 11 requirements. |
Korea (KCC): Class A Equipment (Industrial Broadcasting & Communication Equipment) |
Class A |
This product meets the requirements of industrial electromagnetic equipment, vendors or users should be aware of this. This device is intended for use in a commercial environment, instead of the home environment. |
USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per |